Installation and use of PP folding filter

2019年12月24日

  
1, filter installation method .
 
1. Moisten the O-ring and slowly insert the filter element vertically, all of which must be inserted into the stainless steel round groove. This article introduces the installation and use of PP folding filter element.
 
2. Press the stainless steel orifice plate of the filter element fins, and the pressure plate does not need to be too tight to prevent the filter element from deforming during high temperature disinfection.
 
3. Avoid direct contact with the filter element by hand.
 
4. Rinse the filter element as much as possible before use.
 
5. When turning on or off, slowly turn the valve, do not open or close at one time, in order to prevent the filter from being sucked during high temperature disinfection.
 
2, how to use ?
 
1) Filter element disinfection method
 
1. Disinfection in the disinfection cabinet, remove the filter element from the plastic bag, and put it in the disinfection cabinet at 121 degrees for 30 minutes.
 
2. Please sterilize online, the filter element should be installed in the filter according to the correct method (the distance between the fixed plate and the filter element is 0.5mm). Steam for 30 minutes.
 
 
2) Incoming and outgoing flow direction identification
 
The outside of the filter element enters the middle and exits, and the forward and backward recoil can be carried out in different directions.
 
 
3) Aperture recognition
 
The filter housing has a hot-melt font, which indicates the filter material and aperture.